Seitenansicht 1
©2013 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc. Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com
DIFFERENTIAL PRESSURE
Model 230
Wet-to-Wet Pressure Transducer
DIMENSIONS (Model 230)
3.05
77
2.44
62
Bleed
Screws
Transducer
Mounting Holes
(Screws and mounting bracket supplied)
2.15
55
2.00
51
1.50
38
2.50
64
3.00
76
0.52
13
0.88
22
1.30
33
Dia. 0.156
Dia. 4.0
Notched
for Clamp
Mounting
1/4 ''- 18
NPT high
P
ressure Port
1.94
49
1/4 - 18 NPT Low
Pressure Port
Dia. 0.875 Conduit
Dia. 22 Opening
1.63
41
1.63
41
1.00
25
IN
MM
0.43
11
MOUNTING BRACKET
FR
ONT
VIE
W
SIDE
VIE
W
SPECIFICATIONS
Performance Data Physical Description (Model 230) Electrical Data (Voltage)
Accuracy RSS
1
(at constant temp) ±0.25% FS Case Stainless Steel/Aluminum Circuit 3-Wire (Exc, Out, Com)
Non-Linearity, BFSL ±0.20% FS
Electrical Connection Barrier strip terminal block with conduit enclosure & 0.875
DIA conduit opening.
Excitation 9 to 30 VDC for 0-5 VDC Output
13 to 30 VDC for 0-10 VDC Output
Hysteresis 0.10% FS Pressure Fittings 1/4”-18 NPT internal Output
7
0 to 5 VDC
8
, 0 to 10 VDC
8
Non-Repeatability 0.05% FS Weight (approx.) 14.4 oz Output Impedance 100 ohms
Thermal Eects
2
Sensor Cavity Volume 0.27 in
3
Positive Port, 0.08 in
3
Negative Port
Electrical Data (Current)
Compensated Range °F(°C) +30 to +150 (-1 to +65)
(With 1/4”NPT external ttings installed-does not include cavity volume of 1/4”NPT external ttings.) Circuit 2-Wire
Zero Shift %FS/100°F(%FS/50°C) 2.0 (1.8)
Physical Description (3-Valve Manifold Assembly)
4
Output
9
4 to 20mA
10
Span Shift %FS/100°F(%FS/50°C) 2.0 (1.8) Manifold Block Brass External Load 0 to 1000 ohms
Line Pressure Eect
Zero shift ±0.004% FS/psig
line pressure
Valves (3)
5
V1 for Connection to + port
V2 for Connection to - port V3 for Equalizing Pressure
Minimum supply voltage
(VDC)
9+ 0.02 x (Resistance of receiver plus line).
Resolution
Innite, limited only by
output noise level (0.02%FS)
Valve Type 90° On/O Maximum supply voltage
(VDC)
30+ 0.004 x (Resistance of receiver plus
line).
Static Acceleration Eect 2%FS/g (most sensitive axis)
Process Connections 1/4”-18 NPT Internal Thread
Pressure Media
Natural Frequency 500 Hz (gaseous media)
Dimensions 7.05”W x 6.25”H x 2.16’’D Model 230
Warm-up Shift ±0.1% FS total Weight <2.5 lbs. Gases or liquids compatible with 17-4 PH Stainless Steel, 300 Series
Stainless Steel, Viton and Silicone O-Rings.
Note: Hydrogen not recommended for use with 17-4 PH stainless steel.
Optional Buna-N O’rings are recommended for hydrocarbon applications.
Response Time 30 to 50 milliseconds
Physical Description (5-Valve Manifold Assembly)
6
Long Term Stability 0.5%FS/1 YR Manifold Block Brass
Maximum Line Pressure 350 psig
Valve (5)
5
V1 for Connection to ± Port
V2 for Connection to – Port
V3 for Equalizing Pressure
V4 & V5 for Connection to External
Gauge or Alternate Plumbing Conguration
3 & 5 Valve Manifold
Environmental Data
Gases or liquids compatible with 360 brass, Copper 122, Acetal plug
valves and Nitrile O-rings.
Operating
3
Temperature °F (°C) 0 to +175 (-18 to +80) Process Connection 1/4”-18 NPT Internal Thread
5
Refer to drawings on page16 and 17.
6
Order assembled with the Model 230 (Code 5V)
7
Calibrated into a 50K ohm load, operable into a 5000 ohm
load or greater.
8
Zero output factory set to within ±25mV (for 5 VDC output)
or ±50mV (for 10 VDC output)
Span (Full Scale) output factory set to ±25 mV (for 5 VDC
output ) or ± 50 mV (for 10 VDC output
9
Calibrated at factory with a 24 VDC loop supply voltage and
a 250 ohm load.
10
Zero output factory set to within ±0.16mA. Span factory set
tp wothin ±-.16 mA
Specications subject to change without notice.
Storage Temperature °F (°C) -65 to +250 (-54 to +121) Dimensions 7.05”W x 6.25”H x 2.16’’D
Vibration 5 g from 5 Hz to 500 Hz Weight <3.8 lbs.
Acceleration 10 g
1
RSS of Non-Linearity, Hysteresis, and Non-Repeatability.
2
Units calibrated at nominal 70˚F. Maximum thermal error computed from this datum.
3
Operating temperature limits of the electronics only. Pressure media temperatures
may be considerably higher.
4
Order assembled with the Model 230 (Code 3V) or separately as Option 891.
Shock 50 g
Seitenansicht 1